The flexible detector: ZETALIF EvolutionTM |
A Sensitive solution for HPLC, fast-LC and CE systems |
High Sensitivity: 10 times to 100 times greater
than conventional fluorescence and up to 100,000 times more
sensitive than UV detection.
|
Versatility: 14 interchangelable different lasers availables for any application |
Modularity : Adaptable to CE and HPLC systems
including Fast-LC, Capillary LC and nanoLC. |
Selectivity: Low noise and interference molecules will be detected. Only labeled or native fluorescente |
| Download the Brochure |
ZETALIF EVOLUTION Detector Specifications: |
Capillary :
OD: 365 µm to 460 µm
ID : Refer to Table1 |
| Capillary ID ( µm) |
Estimated*
cell Volume nL |
25 |
0.2 |
50 |
0.4 |
75 |
0.9 |
100 |
1.6 |
150 |
3.5 |
320 |
16.1 |
Table1 : Capillary ID and Volume of Cell
*Based on 200 µm window length
|
|
Two models are available :
LIF-UV-01 Deseigned for lasers emitting at wavelengths from 266 to 488 nm
LIF-VI-01 Deseigned for lasers emitting at wavelengths from 410 to 780 nm
|
Signal outputs: Analog (processed) 0-1V (DC), 0-100 mV (DC)
Analog (direct) 0-1V (DC) |
Available Lasers :
Laser |
wavelength nm |
Source |
Power |
266 |
Pulsed DPSS |
2mW |
325 |
Helium/Cadmium |
10mW |
355 |
Pulsed DPSS |
2mW |
410 |
Diode |
20mW |
473 |
Diode |
30mW |
488 |
Diode |
25mW |
514 |
Argon ion |
40mW |
532 |
DPSS CW |
15mW |
594 |
Helium/Neon |
2mW |
633 |
Helium/Neon |
10mW |
635 |
Laser Diode CW |
20mW |
650 |
Laser Diode CW |
15mW |
780 |
Laser Diode CW |
40mW |
|
Filter block:
The Filter block depend on the laser and contain the appropriate excitation and emission filter.
ZETALIF Evolution detectors are adapted for use of
a broad variety of lasers. The filter block depends on the laser and contains the appropriate excitation and emission filters.
Contact Picometrics for more information |
Power requirements for the Detector:
100/120 VAC, 50/60 Hz, 0.63 A;
220/240 VAC, 50/60 Hz, 0.315 A |
Data Acquisition Systems:
ZETALIF is compatible with any data acquisition system featuring an analog input (0-1 V).
If no analog input is available, an A/D converter is necessary. |
Interface with HPLC system:
Zetalif Evolution is compatible with all LC systems and optimal sensitivity is acheived at flow rates < 0.5 mL/min. Lower sensitivity may be observed at higher flow rates. |
Interface with CE system:
|
Agilent CE ® : External Detector Adapter kit, MS cassette, and an A/D converter (HP.35900E) needed. Capillary length total 55cm/effective 37cm. The system works with the HP Chemstation software. |
Dimensions and weight: 34.0 (H) x 17.0 (W) x 34.0 (D)cm/12,5 kg |
Beckman P/ACE MDQ ®: External Detector Adapter kit, MS cassette and an A/D converter (SS420-X)needed. Capillary length: total 53cm/effective 35cm.The system works with the 32 karat software. |
|
PrinCE® : For single lift models, capillary length: total 70cm/effective 50cm. (80/35 for double lift models). The system works with the DAX software.
Other systems: please contact Picometrics. |
|