ZETALIFTM LED |
|
Request for a quote > |
New LED Induced Fluorescence Detector
A Sensitive solution for HPLC, fast-LC and CE systems
On-line CE-LIF-MS
On-line LC-LIF-MS
|
|

Download the Brochure |
High Sensitivity: 10 times to 100 times greater
than conventional fluorescence and up to 100,000 times more
sensitive than UV detection. |
Modularity : Adaptable to CE and HPLC systems
including Fast-LC, Capillary LC and nanoLC. |
Selectivity: Low noise and interference molecules will be detected. Only labeled or native fluorescente |
| ZETALIF Discovery Detector Specifications: |
Capillary :
OD: 365 µm to 460 µm. ID : Refer to Table1
| Capillary IDµm |
Estimated*
cell Volume nL |
25 |
0.2 |
50 |
0.4 |
75 |
0.9 |
100 |
1.6 |
150 |
3.5 |
320 |
16.1 |
Table1 : Capillary ID and Volume cell. *Based on 200 µm length of window
AvailableWavelengths :
|
| |

On-line CE LIF MS |
|
| |
|
Signal outputs: Analog (processed) 0-1V (DC), 0-100 mV (DC)
Analog (direct) 0-1V (DC)
Distances:
Detector / Cell : 1m (maximum) ;
Dimensions and weight: 34.0(H) x 17.0(W) x 34.0(D)cm/12,5 kg |
Interface with HPLC system:
ZETALIF LED is compatible with all LC systems but optimal sensitivity is acheived at flow rates < 0.5 mL/min. Lower sensitivity may be observed at higher flow rates. |
|
Filter block:
The Emission Filter block generally contain Highpass filter. For specific filter, ask Picometrics.
Zetalif LED detectors are adapted for the use with one wavelength only. .
Contact Picometrics for more information |
|
Power requirements for the Detector:
100/120 VAC, 50/60 Hz, 0.63 A;
220/240 VAC, 50/60 Hz, 0.315 A |
|
Data Acquisition Systems:
ZETALIF is compatible with any data acquisition system featuring an analog input (0-1 V).
If no analog input is available, an A/D converter is necessary. |
|
Interface with CE system:
Agilent CE: Internal Detector Adapter kit, UV cassette, and A/D converter (HP.35900E)
are needed. Minimum capillary length total 33 cm/effective 14 cm.
The system works with the ChemStation software.
Beckman P/ACE MDQ: External Detector Adapter kit, MS cassette and an A/D converter (SS420-X)needed. Capillary length: total 53cm/effective 35cm.The system works with the 32 karat software.
Other systems: please contact Picometrics. |
| |
|